Test Kit is for In-vitro diagnostic (IVD) use to determine Urine albumin (U-Albumin) quantitatively from human urine. It is useful to screen early kidney disease and response to treatment with a chronic disease like diabetes. 


  • Screening early kidney disease
  • Immuno-assay (flow-through)
  • High-throughput more than 60 tests/hour

Specifications

Simple Procedure

  Principle
Flow-through immunometric assay
  Sample Type
Urine
  Sample Volume
 10 𝛍L
  Test Result
 U-Albumin
  Test Range
 3.5 - 200 mg/L
  Precision
CV < 5%
  Test Time
Less than 4 minutes
  Calibration 

Olympus AU 5800,

ELITECH Selectra ProM

Performance

Precision
Within-run
Day-to-day
Lv.1
Lv.2
Lv.3
Lv.1
Lv.2
Lv.3
U-Alb (mg/L)
6.56
51.95
103.76
13.16
52.62
103.89
SD (mg/L)
0.12
1.17
1.67
0.32
0.68
1.27
CV (%)
1.77
2.25
1.61
2.47
1.30
1.23

Contents

Name
Quantity
Package
Remark
TC/Test Cartridge
50 x 1
Individual pouch
Disposable
R1/Reagent
50 x 1
Zipper bag
Disposable
R2/Conjugates
1
Bottle
For 50 tests
R3/Wahing Solution
1
Bottle
For 50 tests
IC/Item Card
1
-
-
IFU/Instructions for use
1--

Test Principle