Test Kit is for In-vitro diagnostic (IVD) use to determine Glycated Hemoglobin (HbA1c) quantitatively from human blood. It is useful at initial assessment and long-term monitoring of glycemic control. 


  • NGSP and IFCC certified. 
  • Boronate Affinity Chromatography
  • Lab-quality precision and accuracy

SPECIFICATIONS

  Assay Principle
 Boronate affinity chromatography
  Sample type
 Whole blood (Capillary, Venous)
  Sample volume
 5 𝛍L
  Test result
HbA1c
  Test range
-  HbA1c : 3 ~ 15 % (NGSP), 9 ~ 140 mmol/mol (IFCC)
- eAG : 39 ~ 384 mg/dL, 2.2 ~ 21.3 mmol/L
  Assay time
 3 minutes
 CV< 3 %
  Reference analyzer
Tosoh G11 (HPLC)
  Storage condition
 - R1, W1 : 2 ~ 10 ℃
- Cartridge : 15 ~ 35 ℃

ORDER INFORMATION 


ContentsCat.NoSpecificationPackaging
AnalyzerEpithod®AutoDx E1M1001 unitStandard carton
Test Kits
Epithod® HbA1c Test Kit
E1C10025 tests / kit
(*) Special cold packing

HbA1c control solutionE1Q100
25 tests, 2 levels(*) Special cold packing
Auxiliaries
Capillary tube
AUX0015 μL, 200 pcs/pack
Standard carton

Safety lancet
AUX20128 gauge, 100 pcs/pack
Standard carton

Pipette tipsAUX301Yellow tip, 1,000 pcs/packStandard carton

Thermal printer paperAUX4015 rolls/packStandard carton

Barcode readerAUX402RQSTStandard carton

External batteryAUX403RQSTStandard carton

(*) Special packing charge will be imposed under MOQ.

HOW TO USE